Menu English Ukrainian russian Home

Free technical library for hobbyists and professionals Free technical library


ENCYCLOPEDIA OF RADIO ELECTRONICS AND ELECTRICAL ENGINEERING
Free library / Schemes of radio-electronic and electrical devices

Household digital thermometer. Encyclopedia of radio electronics and electrical engineering

Free technical library

Encyclopedia of radio electronics and electrical engineering / Power regulators, thermometers, heat stabilizers

Comments on the article Comments on the article

The need for a temperature meter is due to many circumstances. In everyday life, for example, the need to quickly measure the temperature of a person’s body or water, for bathing a child, the temperature inside or outside the room, in a greenhouse or greenhouse, in the basement if vegetables are stored there, in the refrigerator or freezer chamber, water in an aquarium and many others objects.

Household thermometers are usually subject to such requirements as measurement accuracy - no worse than 0,5 C in the temperature range from -50 to +100 ° C (when measuring human body temperature - no worse than 0,1 ... 0,2 ° C), small size , cost-effectiveness, autonomy of power supply, low thermal inertia and hygienic safety. The relatively simple digital thermometer described here largely fulfills these requirements.

The sensitive element of the device is a temperature sensor, the principle of operation of which is based on the property of some materials to change their electrical resistance with a change in temperature. Temperature sensors may be different. In industry, for example, massive metal (copper or platinum) thermal converters are often used.

For household appliances, small-sized semiconductor thermistors MMT, KMT, ST1, ST3, TR-4 are most suitable. MMT-4, which, compared to metal transducers, have much less thermal inertia, have almost ten times higher temperature coefficient of resistance (TCR), higher electrical resistance, which makes it possible to completely neglect the resistance of the wires that connect the sensor to the device. The best characteristics are possessed by a miniature drop-shaped vitrified thermistor TR-4 with a reduced TCR. It has dimensions of 6x4x2,5 mm; flexible leads 80 mm long are made of wire with low thermal conductivity. Its mass is 0,3 g.

The main electrical characteristics of the TR-4 thermistor: nominal resistance - 1 kOhm ± 2% at a temperature of +25 ° C, TKS - approximately 2% / ° C, operating temperature range -60 ... + 200 "C, time constant - 3 s .

Household digital thermometer

The disadvantage of semiconductor thermistors is the non-linearity of the dependence of resistance on temperature and a significant spread in characteristics, which is the main reason that hinders their widespread use for temperature measurement. The graph illustrates the typical dependence of the resistance of TR-4 and MMT-4 semiconductor thermistors on temperature. However, the corresponding circuit design solutions for the linearization of the characteristic can largely eliminate these shortcomings.

The main technical characteristics of the thermometer using the TR-4 thermistor in it:

    Interval of measured temperature, °C. . . -50...+100
    Resolution, °С. . . 0,1
    Measurement error, °С,
    at the edges of the working interval. . . ±0,5
    in the middle part of the working interval, not worse. . . ±0,1...0,2
    Power supply voltage, V. . . 9
    Consumed current, mA. . . one
    Dimensions, mm . . 175x65x30
    Mass, g. . . 250

Schematic diagram of the thermometer is shown in fig. 1. The basis of the device is an integrating analog-to-digital converter (ADC) DA3, to the output of which a four-digit liquid crystal indicator HG1 is connected. Such an element base allowed to reduce power consumption and provide the device with small dimensions and weight.

Household digital thermometer
Fig. 1. Schematic diagram

Household digital thermometer
Rice. 2. PCB

The measuring circuit of the device is formed by a current-setting resistor R1, resistors R2 and R3, which form the reference voltage Uobr, thermistor R4, voltage Ut, on which it changes depending on temperature, and a compensating resistor, the function of which is performed by resistors R5, R6. To reduce the error from self-heating of the thermistor, the value of the current-setting resistor R1 is chosen so that the current in the measuring circuit is approximately 0,1 mA.

The device uses a direct measurement of the thermal resistance by the ratio method - the thermistor R4 and the reference resistor (R2 + R3) are connected in series and the same current flows through them. The voltage drop that occurs across the thermistor is fed to input pins 30 and 31, and the voltage drop across the reference resistor, which acts as a source of reference voltage Uobr, to pins 35 and 36 of the DA3 ADC.

With this method of measurement, the result of the ADC conversion does not depend on the current in the measuring circuit, which means that there is no need for traditionally used high-quality current and reference voltage sources, on which the accuracy characteristics of the meter largely depend.

For a device operating in the temperature measurement mode, a typical problem is to compensate for the initial value of thermal resistance at zero temperature. To do this, the resistance of the compensation resistor (R5 + R6) is chosen equal to the resistance of the thermistor R4 at zero temperature, and in order to compensate for the sum of the voltage values ​​Ut + Uk supplied to the ADC pin 30, a voltage equal to 31 Uk is applied to its pin 2, which forms the operational amplifier DA2 with gain K=(1+R14/R13)=2. Then, taking into account the fact that the resistance of the thermistor decreases with increasing temperature, we have Uin ac = U+in - U-in = 2Uk-(Ut+Uk) = Uk-Ut.

The linearization of the non-linear dependence of the thermal resistance on temperature is implemented by shunting the thermistor R4 with the resistor R11 - roughly, and precisely by introducing the DA1 op-amp into the device. But the shunt resistor R11 only partially rectifies this non-linearity, somewhat expanding the operating temperature range.

The principle of accurate linearization is based on changes in the conversion coefficient of the ADC depending on the reference voltage Uobr. It changes due to feedback through the op-amp DA1. With such a connection, part of the input voltage Uin, determined by the gain of the op-amp DA1 V = [1+(R8+R9)/R7], is added to the voltage Uobr. The more the resistance of the thermistor increases with decreasing temperature, the faster the reference voltage grows, and this leads to a proportional decrease in the ADC conversion coefficient: -arr - voltages at pins 0 and 36 of the ADC, respectively.

If we take the division value of the least significant digit equal to 0,1 C, then in the final form the indication of the digital indicator NG1 will be determined by the expression N=100Uin/Uobr=100(Uk-Ut)/[(U0-B(Uk-Ut)]=100(R5 +R6-R4)/[(R2+R3)-B(R5+R6-R4)]

Other elements of the thermometer that ensure the operation of the ADC are typical. Transistor VT1, turned on by the inverter, serves to indicate the decimal point sign in the digital indicator HG1.

The details of the device are mounted on a printed circuit board made of foil fiberglass with a thickness of 1,5 mm. The DA3 chip is mounted on the side of the printed conductors. Sockets X1, X2 (from the 2PM connector) are soldered directly to the printed circuit boards. Printed pads are also provided for attaching the SA1 switch. Fixed resistors - C2-29V, tuning resistors - SP3-38a. Capacitors: C1 - K50-6, C3 and C7 - K22U, C5 - K73-17, C2 and C6 - K73-24. Switch SA1 - PD9-2, battery GB1 - "Korund". Indicator IZHKTS1-4/8 can be replaced by IZHTS-5.

The design of the sensor is arbitrary. For example, in a plastic rod with a diameter of 5 and a length of 65-70 mm, a through axial hole with a diameter of about 3 mm is drilled, and then a recess is drilled in one of its ends. Thin insulating tubes are put on the leads of the thermistor, the leads are passed into the hole in the rod, the thermistor is installed in the recess and sealed with BOV-1 glue named KO947 varnish. The ends of a two-wire flexible cable are soldered to the conclusions and a piece of thin-walled duralumin tube serving as a sensor handle is tightly put on the end of the rod opposite the thermistor. The length of the connecting cable is about 1,5 m.

Due to the significant variation in the parameters of semiconductor thermistors, three trimming resistors are introduced into the device: R5 - to set zero, R2 - to set the scale scale and R9 - to linearize the thermistor characteristics.

The simplest adjustment of the thermometer is conveniently performed according to three control temperature values: melt water (0 ° C), the human body (36,6 ° C) and boiling water (100 ° C). The first of these test points measures the temperature of the water in the ice, not the water with ice, which may be above 1 °C. At the second control point, a medical thermometer is used as a reference instrument. The boiling point of water must be corrected for atmospheric pressure. In Pyatigorsk, for example, located at an altitude of about 500 m above sea level, water boils at a temperature of 92,5 ° C.

The adjustment is started by placing the sensor in melt water. Trimmer resistor R5 set the indicator to zero. Then, by alternately adjusting the resistors R2 and R9, the indicator readings are obtained corresponding to the temperature values ​​​​at the other two control points. Next, the sensor is again placed in the melt water and all control measurements are repeated.

More precise adjustment of the device can be performed using industrial mercury thermometers with a scale division of 0,2 ° C.

Instead of the TR-4 thermistor, other thermistors of wider application can be used in the sensor, but with the obligatory adjustment of the resistance of some device resistors. So, with MMT-4 with a nominal resistance of 1,3 kOhm, the resistance of the resistor R11 should be reduced to 3,3 kOhm, and with thermistor ST3-19 with a nominal resistance of 2,2 kOhm - up to 3 kOhm.

The operating modes of the ADC when using TR-4 and MMT-4 thermistors in the device are shown in the table. If there are not enough adjustment limits with trimming resistors, except for R11, then you may have to select resistors R3, R6, R8.

Author; V. Suetin, Pyatigorsk; Publication: cxem.net

See other articles Section Power regulators, thermometers, heat stabilizers.

Read and write useful comments on this article.

<< Back

Latest news of science and technology, new electronics:

Machine for thinning flowers in gardens 02.05.2024

In modern agriculture, technological progress is developing aimed at increasing the efficiency of plant care processes. The innovative Florix flower thinning machine was presented in Italy, designed to optimize the harvesting stage. This tool is equipped with mobile arms, allowing it to be easily adapted to the needs of the garden. The operator can adjust the speed of the thin wires by controlling them from the tractor cab using a joystick. This approach significantly increases the efficiency of the flower thinning process, providing the possibility of individual adjustment to the specific conditions of the garden, as well as the variety and type of fruit grown in it. After testing the Florix machine for two years on various types of fruit, the results were very encouraging. Farmers such as Filiberto Montanari, who has used a Florix machine for several years, have reported a significant reduction in the time and labor required to thin flowers. ... >>

Advanced Infrared Microscope 02.05.2024

Microscopes play an important role in scientific research, allowing scientists to delve into structures and processes invisible to the eye. However, various microscopy methods have their limitations, and among them was the limitation of resolution when using the infrared range. But the latest achievements of Japanese researchers from the University of Tokyo open up new prospects for studying the microworld. Scientists from the University of Tokyo have unveiled a new microscope that will revolutionize the capabilities of infrared microscopy. This advanced instrument allows you to see the internal structures of living bacteria with amazing clarity on the nanometer scale. Typically, mid-infrared microscopes are limited by low resolution, but the latest development from Japanese researchers overcomes these limitations. According to scientists, the developed microscope allows creating images with a resolution of up to 120 nanometers, which is 30 times higher than the resolution of traditional microscopes. ... >>

Air trap for insects 01.05.2024

Agriculture is one of the key sectors of the economy, and pest control is an integral part of this process. A team of scientists from the Indian Council of Agricultural Research-Central Potato Research Institute (ICAR-CPRI), Shimla, has come up with an innovative solution to this problem - a wind-powered insect air trap. This device addresses the shortcomings of traditional pest control methods by providing real-time insect population data. The trap is powered entirely by wind energy, making it an environmentally friendly solution that requires no power. Its unique design allows monitoring of both harmful and beneficial insects, providing a complete overview of the population in any agricultural area. “By assessing target pests at the right time, we can take necessary measures to control both pests and diseases,” says Kapil ... >>

Random news from the Archive

Beak repair using a computer 25.08.2015

An American white pelican living in the Dalian Forest Zoo (PRC) has had its damaged beak restored using 3D printing technology.

The male pelican broke his beak in a fight with a competitor during the mating season, as a result of which the bird lost its ability to eat and was doomed to starvation. To save the pet, the staff of the Dalian Forest Zoo turned to the local company Dalian Science and Technology.

Specialists from a Chinese company that is known for 3D printing medical implants examined the beak and decided to remove its damaged part, replacing it with an artificial prosthesis five millimeters thick.

The operation was successful, but as the beak grows, the printed implant will have to be changed. However, the game is worth the candle, as the artificial beak helped the pelican return to good nutrition.

Other interesting news:

▪ superionic ice

▪ The street is lit by the sun and the wind

▪ How to protect your mobile phone from theft

▪ Viscous liquid of electrons flowing in graphene

▪ The battery heals itself

News feed of science and technology, new electronics

 

Interesting materials of the Free Technical Library:

▪ section of the site Videotechnique. Article selection

▪ article Informatics. Lecture notes

▪ Is it right to reduce a dog's behavior to a simple complex of conditioned and unconditioned reflexes? Detailed answer

▪ Ajgon article. Legends, cultivation, methods of application

▪ article Theory: volume and tone control. Encyclopedia of radio electronics and electrical engineering

▪ article VHF prefix to a direct amplification receiver. Encyclopedia of radio electronics and electrical engineering

Leave your comment on this article:

Name:


Email (optional):


A comment:





All languages ​​of this page

Home page | Library | Articles | Website map | Site Reviews

www.diagram.com.ua

www.diagram.com.ua
2000-2024